Publications / Scientific Publications /

Effect of intraocular scattering in macular pigment optical density measurements

Abstract

Fundus reflectometry is a common in-vivo, noninvasive method to estimate the macular pigment optical density (MPOD). The measured density, however, can be affected by the individual’s intraocular scattering. Scattering causes a reduction in the contrast of the fundus image, which in turn leads to an underestimation of the measured density. Intraocular scattering was measured optically in a group of seven young, healthy subjects using the method of optical integration and was subsequently used to correctly estimate the MPOD from fundus images. It was shown that when scattering is not considered, the measured optical density using fundus reflectometry can be underestimated by as high as 16% for our group of subjects

http://dx.doi.org/10.1117/1.JBO.23.5.056001

If you like it, please share it...Tweet about this on Twitter0Share on Facebook0